Series 2600 System SourceMeters Reference Manual Status Model D-21
Return to Appendix D topics 2600S-901-01 Rev. A / May 2006
• Bit B0, Calibrating (CAL) — Set bit indicates that one or more channels
are calibrating.
• Bit B4, Measuring (MEAS) — Bit will be set when taking an overlapped
measurement, but it will not set when taking a normal synchronous
measurement.
• Bit B11, Prompts Enabled (PRMPT) — Set bit indicates that command
prompts are enabled.
• Bit B12, User (USER) — Set bit indicates that an enabled bit in the
operation status user register is set.
• Bit B13, Instrument Summary (INST) — Set bit indicates that an enabled
bit in the operation status instrument summary register is set.
• Bit B14, Program Running (PROG) — Set bit indicates that a program is
running.
Operation Status Calibration Summary
This calibration summary register set feeds to CAL bit B0 of the Operation Status
Register. Used summary bits for this register include B1 (SMUA) and B2 (SMUB).
A set bit indicates that a SMU (A or B) is calibrating.
Operation Status Measurement Summary
This measurement summary register set feeds to MEAS bit B4 of the Operation
Status Register. Used summary bits for this register include B1 (SMUA) and B2
(SMUB). A set bit indicates that a SMU (A or B) is performing an overlapped
measurement.
Operation Status SMU A Summary Register
This SMU summary register set feeds to bit B1 (SMUA) of the Operation Status
Instrument Summary Register. Used summary bits for this register include B0
(CAL), B4 (MEAS), B11 (PRMPT), B12 (USER) and B14 (PROG). A set bit
indicates that the corresponding operation event for SMU A has occurred.
Operation Status SMU B Summary Register
This SMU summary register set feeds to bit B1 (SMUB) of the Operation Status
Instrument Summary Register. Used summary bits for this register include B0
(CAL), B4 (MEAS), B11 (PRMPT), B12 (USER) and B14 (PROG). A set bit
indicates that the corresponding operation event for SMU B has occurred.
Operation Status Instrument Summary Register
This summary register is fed from the SMU summary registers and then feeds to
bit B13 (INST) of the Operation Status Summary Register. Used summary bits for

Table of Contents

Other manuals for Keithley SourceMeter 2612

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2612 and is the answer not in the manual?

Keithley SourceMeter 2612 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2612
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals