8.2 General functions
Menu selection:
Ö Turn on balance using the ON/OFF-button
Ö Press the Tare-button; the display shows horizontal lines
Ö While this display is on, press the Cal-button
Ö Keep F-button pressed until the desired selection is displayed
Function Display
F-button
Selection
PRINT-button
F-button
Description of the
selection options
0
off
Rear illuminated
Display
(see Chapter. 7.1.1)
bl
1
on
AUTO 0*
Auto-zero-
function activated
Auto-zero-function
(see Chapter 6.5.4)
AUTO
AUTO 1
Auto-zero-
function deactivated
Ac_t 0*
Automatic adjustment function
activated
Ac_t
Ac_t 1
Automatic adjustment function
deactivated
AC_c 0*
Automatic adjustment function
activated
Setting the automatic
adjustment function,
only ALJ /PLJ6100-2
in non-verified setting
(see Chapter 6.7.1)
Ac_c
AC_c 1
Automatic adjustment function
deactivated
Pd_d 0*
Last digit is not printed
Data output
(only ALJ / PLJ6100-2
in non-verified setting)
Pd_d
Pd_d 1
Last digit is printed
Vibration filter
Con
Con 1
Con 2
Con 3*
Con 4
Con 5°
Sensitive and fast
(very quiet set-up location)
Robust but slow (very busy
set-up location)
Forming of an
average value
(Animal weighing
program)
AuE
AuE 1
AuE 2*
AuE 3
Fastest display
Slowest display
FIL 0*
Setting for dosing
deactivated
Display speed
FIL
FIL 1
Setting for dosing
activated
Standard weighing
unit
St_uni
see Chapter 7.2.2
* = Factory setting
° = Recommended counter setting