8.3 Parameter for the serial interface
Menu selection:
Ö Turn on balance using the ON/OFF-button
Ö Press the Tare-button; the display shows horizontal lines
Ö While this display is on, press the PRINT -button
Ö Keep F-button pressed until the desired selection is displayed
Function Display
F-button
Choice
PRINT-
button
Description of the
selection options
bod 1
300 bps
bod 2
600 bps
bod 3
1200 bps
bod 4
2400 bps
bod 5*
4800 bps
Baud rate
bod
bod 6
9600 bps
StAb 0*
Output for stable
Weighing value
StAb
(only ALJ
120/220 in
non-
verified
setting)
StAb 1
Output also for
instable weighing value
cont 0*
Consecutive serial output deactivated, output
only when button is pressed
cont
cont 1
Consecutive serial output
dAtE 0*
Printout of the weighing values without
placeholder for date/time
Output
conditions
at the
interface
dAtE
dAtE 1
Printout of the weighing values with
placeholder for date/time
Repl 0
Issue takes place after pressing the PRINT
key.
Repl
(Condition:
cont 0)
Repl 1
Automatic issue of first stable weighing value
Sequence of operations:
1. Tare
2. Place weight, issue of first stable weighing
value
3. No next issue until weight is removed
Condition: Display +/- 50 display steps
from zero
4. Place next weight
* = Factory setting
Ö Confirm setting by pressing the PRINT-button
PLS/PLJ/ALS/ALJ-BA-e-0819
63