11 Troubleshooting
The balance should be switched off for a short time following an interruption in the
programme sequence and disconnected from the mains supply. It is then necessary
to repeat the weighing process from the beginning.
Help:
Fault Possible cause
Weight display is not
illuminated.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
Batteries are inserted incorrectly or empty
There are no batteries inserted.
The weight display changes
continually
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign matter.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously wrong
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Error messages
Zero
Weight too small
FuLL-1
A/D transducer range exceeded
FuLL-2
Maximum load of the balance exceeded
Err-5
Error of the temperature sensor, turn balance
on and off
Err-3
Error for internal adjustment
CAL- Err
Adjustment error or wrong test weight,
Repeat adjustment process
Switch the balance off if other error messages should appear and then switch on
again. If error message persists, inform manufacturer.