KERN
PLJ 360-3M PLJ 510-3M PLJ 2100-2M PLJ 4000-2M PLJ 6100-2
Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g
Weighing range (Max) 360 g 510 g 2,100 g 4,000 g 6,100 g
Minimum load (Min) 0,02 g 0,02 g 0,5 g 0,5 g -
Taring range (subtractive) 360 g 510 g 2,100 g 4,000 g 6,100 g
Reproducibility 0.002 g 0.002 g 0.02 g 0.02 g 0.01 g
Linearity ± 0.002 g ± 0.002 g ± 0.02 g ± 0.02 g ± 0.03 g
Minimal piece weight for
Parts counting
> 0.05 g > 0.005 g > 0.05 g > 0.05 g > 0.05 g
Adjustment weight internal internal internal internal internal
Verifiable yes yes yes yes no
Verification value (e) 0.01 g 0.01 g 0.1 g 0.1 g -
Accuracy category II II II II -
Warming up time (operating
temperature)
2 h 4 h 2 h 4 h 4 h
Reference piece numbers for
parts counting
10, 20, 50, freely selectable
Weighing Units mg, g, ct,
mg, g, ct, gn,
mom, oz, dwt
Stabilization time (typically) 4 sec.
Permissible ambient temperature + 15° C .... +30° C
Humidity of air max. 80 % (not condensing)
Under floor weighing device Loop for under floor weighing, serial
Weighing plate mm 128 x 128 165 x 165
Housing (W x D x H) mm 206 x 312 x 160
(with draft shield)
206 x 312 x 98
(without draft shield)
Weight kg (net) 4,6 5,4
PLS/PLJ/ALS/ALJ-BA-e-0819
40