5 Replaceable Parts
124 Keysight E4416A/E4417A Power Meters Service Guide
Replacing the Rear Panel Power Reference Cable Assembly
(Option 003)
1 Disassemble the power meter down to allow access to the rear panel
connectors. (Refer to “Removing the A4 Comms Assembly” on page 111 for
further information).
2 Replace the power reference cable assembly using the replacement kit and the
torques indicated on the following diagrams:
25 lb/in
6 lb/in
tie wraps
9 lb/in