EasyManuals Logo

Keysight Technologies E4417A User Manual

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 31
Instrument Accuracy (Normal Path) Test
Specification
Description
This test verifies the accuracy of the normal path in the power meter by
calculating linear error over 50 samples. The standard deviation is then calculated
using the non-biased or n-1 method:
Equation 2-1: Standard Deviation
The test result is obtained by applying a factor of K = 2.58:
Equation 2-2: Accuracy
This accuracy figure represents electrical performance of the normal path with
confidence limits of 99%.
Electrical Characteristics Performance Limits (maximum)
Accuracy 0.8% or 0.04 dB
σ
nΣx
2
Σx()
2
nn 1()
--------------------------
=
Accuracy 2.58
σ
=

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals