2 Performance Tests
42 Keysight E4416A/E4417A Power Meters Service Guide
Test Procedure 2
Equipment
Range Calibrator: Keysight 11683A Option H01
Power Sensor Cable: Keysight E9288A
DC Calibrator: Datron 4000A
NOTE
This test procedure makes use of a 11683A Range Calibrator fitted with Option
H01. The measurement uncertainty of this test procedure is significantly lower
than the previously described method (refer to Appendix D, Measuremet
Uncertainty Analysis - Instrument Accuracy Test for a description of the
measurement uncertainty analysis). As such, this test procedure should be used
to ensure the power meter meets its published specification.