Replaceable Parts 5
Keysight E4416A/E4417A Power Meters Service Guide 105
Front Panel Connector Repair
The front panel assembly is an exchange assembly. However, if front panel sensor
cables or power reference cables are faulty they should be replaced by ordering
the appropriate kit and following the assembly/disassembly procedure. The
instrument accuracy tests should then be carried out to verify the functionality of
the new part. Refer to Chapter 2 for further information on instrument accuracy
tests.