Adjustments 3
Keysight E4416A/E4417A Power Meters Service Guide 73
Post-Repair Adjustments
Table 3-1 lists the adjustments related to repairs or replacement of any of the
assemblies.
Table 3-1 Post Repair Adjustments, Tests, and Checks
Assembly Replaced Related Adjustments, Performance
Tests or Self Tests
A1 Power Supply Self Test
A2 Processor Assembly Power Reference Frequency and Level Test/Adjustment.
Display Brightness and Contrast Adjustment.
A3 Front Panel Assembly Instrument Accuracy Test, Display
Brightness and Contrast Adjustment.
A4 Comms Assembly All automated Tests and Serial Interface
Self Test.
A5 Daughter Assembly Instrument Accuracy Test.
A6 Measurement Assembly Instrument Accuracy Test.
W1 or W2 Power Sensor Cable Instrument Accuracy Test.
BT1 Lithium Battery Battery Self Test.
Power Reference Cable Kits Power Reference Level Test.