6 Troubleshooting
128 Keysight E4416A/E4417A Power Meters Service Guide
Introduction
This chapter enables qualified service personnel to diagnose suspected faults
with the power meter Rmt I/O (Remote Input/Output) signal lines and RS232/422
serial port.
If there is a problem when attempting to use the RS232/422 serial interface or the
remote I/O functions, consult the User's Guide and confirm that all the user
setups are correct before proceeding with the following fault finding flow charts.
Suggested Diagnostic Equipment
1 Digital multi-meter capable of measuring voltage and resistance, for example,
Keysight E4975.
2 Signal source capable of producing 300 ms TTL single shot pulses, for
example, Keysight 33120A.
3 +5 V TTL logic level source.
4 RS232 self test connector −9 way 'D' type female connector with the following
pins wired together:
– Pin 2 (Rx)to Pin 3 (Tx).
– Pin 4 (DTR)to Pin 6 (DSR).
– Pin 7 (RTS)to Pin 8 (CTS).
5 RS422 self test connector - 9 way 'D' type female connector with the following
pins wired together:
–Pin 1 (CTS−)to Pin 9 (RTS−).
– Pin2 (Rx−)to Pin 4 (Tx−).
– Pin 3 (Tx+)to Pin6 (Rx+).
– Pin7 (RTS+)to Pin 8 (CTS+).