EasyManuals Logo

Keysight Technologies E4417A User Manual

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
6 Troubleshooting
136 Keysight E4416A/E4417A Power Meters Service Guide
Figure 6-7 Troubleshooting Guide - TTL Inputs (continued)
B
Remove A2 Processor Assy (see chapter 5) to gain access
to the Communications Assembly connector.
Measure resistance between Rmt I/O connector pin 6
and connector J4 pin 43.
Meter reading
470Ω +/− 10%
Measure resistance between Rmt
I/O connector pin 5 and connector
J4 pin 44.
Meter reading
470Ω +/− 10%
A2 Processor assy defective
End
A4 Comms assy defective
No
No
Yes
Yes

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals