Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 29
Test Setup
Figure 2-1 Zero Test Setup
Procedure
The following procedure should be performed for the Zero Test.
1 Connect the equipment as shown in Figure 2-1.
2 Switch the power meter on.
3 Press then .
4 Press , , for a reading in watts.
5 Set the range calibrator as follows:
–RANGE: 3μW
POLARITY: NORMAL
–FUNCTION: STANDBY
–LINE: ON
Keysight 11730A
Power
Meter
CHANNEL A
Power Meter
Range
Calibrator
NOTE
The procedure details the key presses required on the Keysight
Technologies E4418B. For the Keysight Technologies E4419B, the equivalent key
presses should be performed on both channels.
Preset
Local
Confirm
Meas
Display
Units
W