EasyManua.ls Logo

Keysight Technologies E4417A

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replaceable Parts 5
Keysight E4416A/E4417A Power Meters Service Guide 125
Replacing the Rear Panel Recorder Output(s) Cable Assembly
1 Disassemble the power meter down to allow access to the rear panel
connectors. (Refer to “Removing the A4 Comms Assembly” on page 111 for
further information).
2 Replace the recorder output cable assembly using the replacement kit and the
torque indicated on the following diagram:
25 lb/in
NOTE
Recorder output A connects to A4 J23 and recorder output B connects to
A4 J24.

Table of Contents

Other manuals for Keysight Technologies E4417A

Related product manuals