Replaceable Parts 5
Keysight E4416A/E4417A Power Meters Service Guide 125
Replacing the Rear Panel Recorder Output(s) Cable Assembly
1 Disassemble the power meter down to allow access to the rear panel
connectors. (Refer to “Removing the A4 Comms Assembly” on page 111 for
further information).
2 Replace the recorder output cable assembly using the replacement kit and the
torque indicated on the following diagram:
25 lb/in
NOTE
Recorder output A connects to A4 J23 and recorder output B connects to
A4 J24.