EasyManua.ls Logo

Keysight Technologies E4417A

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C Measurement Uncertainty Analysis - Power Reference Level Test
152 Keysight E4416A/E4417A Power Meters Service Guide
with;
0.15 ppm of Reading + 0.01 ppm of Range per
o
C (temp coefficient Tcal ±5
o
C)
± 8.75 ppm of Reading + 0.55 ppm of Range
Therefore,
For V
0
and V
1
, two measurements are taken, so the uncertainty is equated to the
RSS of using two DMMs;
UV
comp
()UV
ref
()
8.75 5×
1000000
-----------------
0.55 10×
1000000
------------------+± 49.25
μ
V±== =
Voltage measurement Uncertainty
U (V
comp
)
49.25 μV
U (V
o
)
69.65 μV
U (V
1
)
69.65 μV

Table of Contents

Other manuals for Keysight Technologies E4417A

Related product manuals