Measurement Uncertainty Analysis - Power Reference Level Test C
Keysight E4416A/E4417A Power Meters Service Guide 157
The readings in Table C-1 are in milli-Watts, therefore the Standard Deviation
(worse case) is: - U (Rep) = 0.057 mWatts
14 0.999513893 0.999553753
15 0.999589288 0.999496811
16 0.999485404 0.999529486
17 0.999559381 0.999534985
18 0.999524217 0.99948097
19 0.999523013 0.999518899
20 0.999534276 0.999488886
Max 0.999647 0.999566084
Min 0.999406 0.999377721
Average 0.999529 0.999504053
SD 5.67E-05 5.01868E-05
[a] Measurement 5 was discounted from the calculations as being an extraneous measurement.
Table C-1 Measurement Repeatability Results (continued)
Measurement ADJUST REF. REF. LEVEL TEST