2 Performance Tests
34 Keysight E4416A/E4417A Power Meters Service Guide
7 Turn on the power reference:
Press
,
8 Connect the E9321A sensor to the power reference.
9 Calibrate the system:
For the E4416A press
,
For the E4417A press ,
,
10 Provide a point of reference against which the power meter can base its
measurements:
Setup: press
, , , ,
,
,
Entry: read the value shown on the DVM display. Set the reference voltage
on the pop-up window on the power meter to this value and press .
This value provides a point of reference that allows the power meter and
DVM measurements to be compared at varying power levels.
Zero
Cal
Power Ref On
Zero
Cal
Cal
Zero
Cal
Cal
Cal A
System
More
Service
Self Test
More
Fast Path Accuracy
Ref Volts at 0.0 dBm
Enter
NOTE
When operating the power meter in this test mode, the range of power levels
that can be applied is +3 dBm to 30 dBm. Power levels outside this range may
cause inaccurate results to be obtained and in some cases may cause damage
to the E9321A sensor.

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals