Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 65
21 Calculate the power measured from the power reference oscillator using
equation 2-9:
Equation 2-9:
22 Record the value P
1
:
–P
1
___________
23 Calculate the power measured with mount resistance set to 100 Ω:
Set the Keysight 432A mount resistance switch to 100 Ω.
Connect the equipment as shown in Figure 2-10 ensuring that the DVM
input leads are isolated from chassis ground.
Repeat steps 15 to 20. Record the new values of V
0
, V
1
and V
comp
:
–V
0
___________
–V
1
___________
–V
comp
___________
24 Calculate the power measured from the power reference oscillator using
Equation 2-10 and record the value:
Equation 2-10:
–P
2
___________
P
1
2 V
comp
V
1
V
0
()V
0
2
V
1
2
+
4 R
1
----------------- ---------------------------- --------------=
P
2
2 V
comp
V
1
V
0
()V
0
2
V
1
2
+
4 R
2
------------------ ---------------------------- -------------=

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals