EasyManuals Logo

Keysight Technologies E4417A User Manual

Keysight Technologies E4417A
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Performance Tests 2
Keysight E4416A/E4417A Power Meters Service Guide 67
28 Verify that the output SWR is not greater than the maximum value shown in
the following table:
NOTE
The and values are measuredfor the thermistor mount (478A) using the
network analyzer. For clarification, above is measured at 200 ohm while
is measured at 100 Ω.
Γ
1
Γ
2
Γ
1
Γ
2
Table 2-8 Output SWR Result
Max Actual Result
1.06
[a]
__________________
1.08
[b]
__________________
[a] For instruments with the 50 MHz 1 mW output (power ref) connector mounted on the front panel.
[b] For instruments with the 50 MHz 1 mW output (power ref) connector mounted on the rear panel. This applies to models E4416A and E4417A
with Option 003.

Table of Contents

Other manuals for Keysight Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4417A and is the answer not in the manual?

Keysight Technologies E4417A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals