Keysight E5061B Network Analyzers 105
Post-Repair Procedures
Post-Repair Procedures
A2 Receiver
Module
Perform the following required
adjustments using “A2 Receiver Module”
in Spot Adjustment of the program.
“To Execute the Diagnostic Test” on
page 33
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
RF Output Level Linearity Test
Trace Noise Test
Crosstalk & System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
Input Impedance Test (Opt 3L5)
Absolute Amplitude Accuracy Test (Opt 3L5)
Magnitude Ratio/Phase Frequency
Response Test (Opt 3L5)
Trace Noise Test (Opt 3L5)
Crosstalk and Noise Level Test (Opt 3L5)
Magnitude Ratio/Phase Dynamic Accuracy
Test (Opt 3L5)
DC Measurement Accuracy Test (Opt 3L5)
Table 5-1 Post-Repair Procedures (Option 3L5)
Replaced
Assembly or Part
Required Adjustments
Correction Constants (CC)
Verification