106 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimizing Performance for All Models
Using Flatness Correction
The UF annunciator appears in the AMPLITUDE area of the display, and the correction data in the
array is applied to the RF output.
Recalling and Applying a User Flatness Correction Array
The following example assumes that a user flatness correction array has been created and stored. If
not, perform the Example: A 500 MHz to 1 GHz Flatness Correction Array with 10 Correction
Values on page 103.
1. Preset the signal generator.
2. Recall the desired User Flatness Correction file:
a. Press AMPTD > More > User Flatness > Configure Cal Array > More > Preset List > Confirm
Preset.
b. Press Load/Store.
c. Highlight the desired file.
d. Populate the user flatness correction array with the data contained in the selected file:
Press Load From Selected File > Confirm Load From File.
The user flatness correction array title displays User Flatness: Name of File.
3. Apply the correction data in the array to the RF output: Press Return > Flatness Off On to On.

Table of Contents

Other manuals for Keysight Technologies N5182B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5182B MXG and is the answer not in the manual?

Keysight Technologies N5182B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5182B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals