Keysight EXG and MXG X-Series Signal Generators User’s Guide 405
Using BERT for N5172B/82B with Option UN7
Bit Error Rate Test
Clock Delay Function
In this example, the clock delay function is off. Figure 16-4 shows the input of the internal error
detector of UN7 through AUX I/O and indicates that the data is delayed from the clock.
Figure 16-4
CH1: BER TEST OUT (pin 17 of AUX I/O connector)
CH2: BER MEAS END (pin 15 of AUX I/O connector)
In this example, the clock delay function is on. The rising edge of the clock was delayed by 200 ns
and was adjusted to the center of the data. Figure 16-5 indicates the result of the using the clock
delay function.
Figure 16-5
CH1
CH2
CH1
CH2