Keysight EXG and MXG X-Series Signal Generators User’s Guide 411
Using BERT for N5172B/82B with Option UN7
Bit Error Rate Test
Testing Signal Definitions
The timing diagram <Keysight Red >Figure 16-13, “Testing Signal Definitions,” shows the
relationships between a trigger event and the output signals at the BER MEAS END and BER TEST
OUT connectors.
If a BER MEAS END signal stays high following a trigger event, the BERT measurement is in
progress and other trigger events are ignored. This state is stored in the status register and can be
queried.
Figure 16-13 Testing Signal Definitions
T1 is a firmware handling time measured from a Trigger event to the rising edge of a BER MEAS
END signal.
T2 is a firmware handling time measured from the falling edge of a BER TEST OUT signal to the
falling edge of the BER MEAS END signal.
T3 is a minimum requirement time measured from the falling edge of the BER MEAS END signal
to the next trigger event. T3 should be greater than 0 second.
The pulse output of the BER TEST OUT for the Nth-1 test result ends prior to the falling edge of the
BER MEAS END signal for the Nth measurement; so you can use this edge to start latching the Nth
test result.

Table of Contents

Other manuals for Keysight Technologies N5182B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5182B MXG and is the answer not in the manual?

Keysight Technologies N5182B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5182B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals