LAN Interface as-shipped settings
These settings are optimized for connecting your power analyzer to a site network. They should also work well for other network
configurations.
The LANas-shipped settings can be restored by selecting the System\IO\LAN\Defaults menu.
Get IP Address Automatic
Subnetmask 255.255.0.0
Default gateway 0.0.0.0
Host name Keysight-<serial number>
mDNS service name Keysight N67xxx DC Power Analyzer <serial number>
LANservice - VXI-11 Enabled
LANservice - Telnet Enabled
LANservice - mDNS Enabled
LANservice - Web server Enabled
LANservice - sockets Enabled
Web password Blank
Initial N678xA SMU Emulation Mode Settings
The following tables document the emulation mode settings that are applied when the emulation
mode is selected. Emulation mode settings are not saved. N/A means not applicable.
4 Quadrant Power Supply
1
2 Quadrant Power Supply 1 Quadrant Power Supply
Priority mode voltage current voltage current voltage current
Quadrants 4 4 2 2 1 1
Voltage range 6.12 V N/A 6.12 V N/A 6.12 V N/A
Voltage 0 V N/A 0 V N/A 0 V N/A
I limit range 3.06 A N/A 3.06 A
3
N/A 3.06 A
3
N/A
+ I limit 3.06 A N/A 3.06 A
3
N/A min. N/A
I limit 3.06 A N/A 3.06 A
3
N/A 0.612 A N/A
Current range N/A 3.06 A N/A 3.06 A
3
N/A 3.06 A
3
Current N/A 0 A N/A 0 A N/A 0 A
V limitrange N/A 6.12 V N/A 6.12 V N/A 6.12 V