Front Panel Menu Reference SCPI Command
From the Calibration menu, selectMiscellaneous, Downprogrammer. Select
Start. Then selectClose.
CAL:DPRog (@1)
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
From the Calibration menu, selectDate. Enter the date, then selectInsert Date.
Then selectClose.
From the Calibration menu, selectSave. SelectSave Calibration for all Outputs.
Then selectClose.
From the Administrative Tools menu, select Administrator Login/Logout to log
out of the Admin menu.
To enter a calibration date:
CAL:DATE "<date>"
To save the calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
8 Verification and Calibration
398 Keysight N6705C Operating and Service Guide