Step 5. Select the third resistance calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P3 measured data”
Enter the data from the external DMM. Press Selectwhen done. Press Next to
finish calibration.
CAL:LEV P3
*OPC?
CAL:DATA <data>
Step 6. For Model N6781A repeat steps 1 through 5 for the 6 V resistance range. This time, select the
Low range.
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
From the Calibration menu, selectDate. Enter the date, then selectInsert Date.
Then selectClose.
From the Calibration menu, selectSave. SelectSave Calibration for all Outputs.
Then selectClose.
From the Administrative Tools menu, select Administrator Login/Logout to log
out of the Admin menu.
To enter a calibration date:
CAL:DATE "<date>"
To save the calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
8 Verification and Calibration
408 Keysight N6705C Operating and Service Guide