Characteristic Keysight N6705C
Digital Control Characteristics
Maximum voltage ratings
Pins 1 and 2 as FLT output
Pins 1 - 7 as digital/trigger
outputs (pin 8 = common)
Pins 1 - 7 as digital/trigger
inputs and pin 3 as INH input
(pin 8 = common)
+16.5 VDC/-5 VDC between pins
(pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 1 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.8 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
InterfaceCapabilities
LXI Core 2011:
USB 2.0 (USB-TMC488):
10/100/1000 LAN:
Built-in Web server:
GPIB:
10/100/1000 Base-T Ethernet(Sockets, VXI-11 protocol, Web interface)
Requires Keysight IO Library version M.01.01 or 14.0 and up
Requires Keysight IO Library version L.01.01 or 14.0 and up
Requires a Web browser
SCPI - 1993, IEEE 488.2 compliant interface
Regulatory Compliance:
EMC:
Safety:
Complies with European EMC Directive for test and measurement products:
IEC/EN 61326-1; CISPR 11, Group 1, class A; AS/NZS CISPR 11; ICES/NMB-001
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE-marking.
Conforms to US and Canadian safety regulations.
Environmental
Operating environment:
Temperature range:
Relative humidity:
Altitude:
Storage temperature:
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C (output current is derated 1% per °C above 4C
Up to 95% (non-condensing)
Up to 2000 meters
-30°C to 70°C
Acoustic Noise Declaration
This statement is provided to
comply with the requirements
of the German Sound Emission
Directive, from 18 January 1991
Sound Pressure Lp <70 dB(A), At Operator Position, Normal Operation, According to EN 27779
(Type Test).
Schalldruckpegel Lp <70 dB(A), Am Arbeitsplatz, Normaler Betrieb, Nach EN 27779 (Typ-
prüfung).
Output Terminal Isolation: No output terminal may be more than 240 VDC from any other terminal or chassis ground.
N6781A/N6785A Note:
When using the AUX measurement input terminals on Models N6781A and N6785A, no output or
input terminal may be more than ±60 VDC from any other terminal or chassis ground.