666 Keysight N9927-90001 User’s Guide
System Settings
Service Diagnostics
——Check the Test Calibration box and then press Run to run the
Self-Test using OSL standards
With either choice the Self-Test results are displayed on the FieldFox
as they are performed. Refer to Figure 29-10 on page 667.
After the Self-Test completes a pass/fail message is displayed and
A Logfile is stored in the FieldFox here:
[Internal]:Self-TestLogFiles\Self-TestLogFileYYYY.MM.
DD.HH.mm.txt