SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
Effective Isotropic Radiated Power (EIRP) – (Requires CPU2, SA Mode Only)
Effective Isotropic Radiated Power (EIRP) – (Requires CPU2, SA
Mode Only)
How to Select Effective Isotropic Radiated Power (EIRP)
—Press Measure 1
—Then Channel Measurements > More > EIRP to open a menu of softkeys to
set up EIRP.
When EIRP is selected, the following settings are available / maintained from a
previous measurement: ???.
How to Setup EIRP (Effective Isotropic Radiated Power)
General setup and an example for EIRP:
For this example we are using an isotropic antenna with the following
specifications:
Frequency range: 9 kHz to 9 GHz isotropic antenna
Preamp stages: 2 (+15 dB each (@2 GHz)
Chop/Switch speed rate: 1 Hz to 50 kHz
Assume: 600 Series 50
coaxial cable with 2.5 dB/100 feet
1. Press Measure 1
> Channel Measurements > Channel Power > More >
EIRP
then choose from the following:
EIRP Meas
– Enables and disables the EIRP measurement.
Rx Ant Gain
– Enter the receiver ant gain (–100 db to 100 db;
default = 0 dB).
For this example: 30 dB > Enter.
Dist to Ant
– Enter the distance between the receiving and
transmitting antennas (0 m to 25 km; default = 100m)
For this example: 100m (default)
Rx Cable Loss
– Enter the receiver cable loss (0 to –100 dB; default
= 0 dB).
For this example: 2.5 dB
Margin Comp
– Enter the margin compensation to remove known
self-interference, antenna efficiency, fading and other errors
(–100 db to 100 db; default = 0 dB).
Then More
and choose from the following: