5-4 Site Preparation
5 Environmental Requirements
Cooling Requirements
Design the site cooling capability using the heat dissipation estimates in Table 5-2.
Humidity Requirements
The system is designed to operate in the range from 5% to 80% relative humidity
(non-condensing).
If the system is subjected to condensation, as if moved from a cold loading dock
into a warm environment, allow at least 24 hours for the system to recover before
powering up.
Table 5-2 System heat dissipation
Number of Modules Dissipation
1 module 10500 BTU (3.1 kW)
2 modules 19700 BTU (5.8 kW)
3 modules 28500 BTU (8.4 kW)
4 modules 3700 BTU (10.9 kW)