Remote Interface Reference 3
Output Settings and Operation Commands
E3632A User’s Guide 85
Output Settings and Operation Commands
This section describes low- level commands used to program 
the power supply. Although the APPLy command provides 
the most straightforward method to program the power 
supply, the low- level output setting commands give you more 
flexibility to change the individual parameters.
CURRent {<current>|MINimum|MAXimum|UP|DOWN}
This command programs the immediate current level of the 
power supply. The immediate level is the current value of 
the output terminals.
The  CURRent command changes the output of the power 
supply to the newly programmed value regardless of the 
output range presently selected.
You can substitute MINimum or MAXimum in place of a 
specific value for the current parameter. MIN selects the 
lowest current values of 0 A. MAX selects the highest current 
values allowed for the selected range.
This command also increases or decreases the immediate 
current level using the UP or DOWN parameter by a 
predetermined amount. The command CURRent:STEP sets 
the amount of increment and decrement. Notice that a new 
increment setting will cause an execution error - 222 (Data 
out of range) when the maximum or the minimum rated 
current is exceeded.
CURRent? [MINimum|MAXimum]
This query returns the presently programmed current level 
of the power supply. CURR? MAX and CURR? MIN return the 
highest and lowest programmable current levels for the 
selected range.