Error Messages 4
Calibration Error Messages
E3632A User’s Guide 137
Calibration Error Messages
The following errors indicate failures that may occur during
a calibration. Refer to the Service Guide for more
information.
Table 4 - 3 Calibration error messages
701 Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the power supply. When
applicable, this error will occur at power-on to warn you that the power supply is unsecured.
702 Cal secured
The power supply is secured against calibration.
703 Invalid secure code
An invalid calibration security code was received when attempting to unsecure or secure the power
supply. You must use the same security code to unsecure the power supply as was used to secure it,
and vice versa. The security code may contain up to 12 alphanumeric characters. The first character
must be a letter.
704 Secure code too long
A security code was received which contained more than 12 characters.
705 Cal aborted
A calibration in progress is aborted when you press any front-panel key, send a device clear, or change
the local/remote state of the instrument.
708 Cal output disabled
Calibration is aborted by sending the OUTP OFF command during an output calibration.
712 Bad DAC cal data
The specified DAC calibration values (CAL:VOLT or CAL:CURR) are out of range. Note that the new
calibration constants are not stored in the non-volatile memory.
713 Bad readback cal data
The specified readback calibration values (CAL:VOLT or CAL:CURR) are out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
714 Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the new calibration constants
are not stored in the non-volatile memory.