4 Error Messages
Calibration Error Messages
138 E3632A User’s Guide
715 Bad OCP cal data
The overcurrent protection calibration constant is out of range. Note that the new calibration constants
are not stored in the non-volatile memory.
716 Bad DAC DNL error correction data
Invalid data measured during the calibration for DAC differential nonlinearity error correction.
717 Cal OVP or OCP status enabled
Overvoltage protection status or overcurrent protection status is enabled. You must set both overvoltage
and overcurrent protection status to OFF before and during the calibration.
740 Cal checksum failed, secure state
741 Cal checksum failed, string data
742 Cal checksum failed, store/recall data in location 0
743 Cal checksum failed, store/recall data in location 1
744 Cal checksum failed, store/recall data in location 2
745 Cal checksum failed, store/recall data in location 3
746 Cal checksum failed, DAC cal constants
747 Cal checksum failed, readback cal constants
748 Cal checksum failed, GPIB address
749 Cal checksum failed, internal data
750 Cal checksum failed, DAC DNL error correction data
Table 4 - 3 Calibration error messages (continued)