Remote Interface Reference 3
Output Settings and Operation Commands
E3632A User’s Guide 87
CURRent:TRIGgered 
{<current>|MINimum|MAXimum}
This command programs the pending triggered current level. 
The pending triggered current level is a stored value that is 
transferred to the output terminals when a trigger occurs. A 
pending triggered level is not affected by subsequent 
CURRent commands.
CURRent:TRIGgered? [MINimum|MAXimum]
This query returns the triggered current level presently 
programmed. If no triggered level is programmed, the 
CURRent level is returned. CURR:TRIG? MAX and 
CURR:TRIG? MIN return the highest and lowest 
programmable triggered current levels.
CURRent:PROTection 
{<current>|MINimum|MAXimum}
This command sets the current level at which the 
overcurrent protection (OCP) circuit will trip. If the peak 
output current exceeds the OCP level, then the output 
current is programmed to zero. The Questionable Status 
register “OC” bit is set (refer to 
“SCPI status system”  on 
page  108). An overcurrent condition can be cleared with the 
CURR:PROT:CLE command after the condition that caused 
the OCP trip is removed.
CURRent:PROTection? {MINimum|MAXimum}
This query returns the overcurrent protection trip level 
presently programmed. CURR:PROT? MAX and CURR:PROT? 
MIN return the maximum and minimum programmable 
overcurrent trip levels.