EasyManuals Logo

Keysight E3632A User Manual

Keysight E3632A
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
Remote Interface Reference 3
System-Related Commands
E3632A User’s Guide 97
OUTPut?
This query returns the output state of the power supply. The
returned value is 0 (OFF) or 1 (ON).
OUTPut:RELay {OFF|ON}
This command sets the state of two TTL signals on the
RS- 232 connector. These signals are intended for use with
an external relay and relay driver. The TTL output is
available on the RS- 232 connector pin 1 and pin 9. When
the OUTPut:RELay state is ON, the TTL output of pin 1 is
high (4.5 V) and pin 9 is low (0.5 V). The levels are reversed
when the OUTPut:RELay state is OFF. At *RST, the
OUTPut:RELay state is OFF.
TTL output of pin 1 or pin 9 of the RS- 232 connector is
available only after installing two jumpers inside the power
supply. Refer to the Service Guide for more information.
OUTPut:RELay?
This command returns the state of the TTL relay logic
signals. See also OUTP:REL command.
SYSTem:BEEPer
Issue a single beep immediately.
NOTE
Do not use the RS-232 interface if you have configured the power supply to
output relay control signals. Internal components on the RS-232 circuitry
may be damaged.

Table of Contents

Other manuals for Keysight E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E3632A and is the answer not in the manual?

Keysight E3632A Specifications

General IconGeneral
BrandKeysight
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals