Remote Interface Reference 3
Calibration Commands
E3632A User’s Guide 105
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of 
the power supply. It takes about seven seconds to execute 
the command. The calibration must be unsecured and the 
output be opened before calibrating the overvoltage 
protection circuit. The power supply automatically performs 
the calibration and stores the new overvoltage constant in 
non- volatile memory. Notice that voltage calibration precedes 
before sending this command.