EasyManua.ls Logo

Keysight E3632A - Page 153

Keysight E3632A
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Messages 4
Execution Error Messages
E3632A User’s Guide 131
-113 Undefined header
A command was received that is not valid for this power supply. You may have misspelled the command
or it may not be a valid command. If you are using the short form of the command, remember that it may
contain up to four letters.
Example: TRIGG:DEL 3
-114
[1]
Header suffix out of range
The numeric suffix attached to a command header is not one of the allowable values.
Example: OUTP2 ON
-120
[1]
Numeric data error
An invalid number was specified for a numeric parameter.
Example: APPL 1.0E+320000
-121 Invalid character in number
An invalid character was found in the number specified for a parameter value.
Example: *ESE #B01010102
-123 Numeric overflow
A numeric parameter was found whose exponent was larger than 32000.
-124 Too many digits
A numeric parameter was found whose mantissa contained more than 255 digits, excluding leading
zeros.
-128 Numeric data not allowed
A numeric parameter was received but a character string was expected.
Example: DISP:TEXT 123
-130
[1]
Suffix error
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix or the
numeric parameter does not accept a suffix.
Example: TRIG:DEL 0.5 SECS
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix.
Example: TRIG:DEL 0.5 SECS
-134 Suffix too long
A suffix for a numeric parameter contained too many characters.
Table 4 - 1 Execution error messages (continued)

Table of Contents

Other manuals for Keysight E3632A

Related product manuals