16 2000/3000 X-Series Oscilloscopes Service Guide
1 Characteristics and Specifications
Environmental Conditions
Environment Indoor use only.
Ambient
temperature
Operating 0 °C to +55 °C; non-operating –40 °C to +71 °C
Humidity Operating: Up to 80% RH at or below +40
°C. Up to 45% RH up to +50 °C.
Non-operating: Up to 95% RH up to +40 °C. Up to 45% RH up to +50 °C.
Altitude Operating and non-operating to 4,000 m (13,123 ft)
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight InfiniiVision 2000 X Series and is the answer not in the manual?

Keysight InfiniiVision 2000 X Series Specifications

General IconGeneral
BrandKeysight
ModelInfiniiVision 2000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals