36 2000/3000 X-Series Oscilloscopes Service Guide
2 Testing Performance
3 Set up the oscilloscope.
a Set the Volts/Div setting to the value in the first line in Table 11 or Table 12
(depending on the oscilloscope model).
b Adjust the channel 1 position knob to place the baseline at 0.5 major division
from the bottom of the display.
Table 11 Settings Used to Verify Dual Cursor Accuracy, 2000 X-Series Models
Volts/Div Setting Oscilloscope Calibrator
Setting
Test Limits
5 V/Div 35 V 33.6 V to 36.4 V
2 V/Div 14 V 13.44 V to 14.56 V
1V/Div 7V 6.72V to 7.28V
500 mV/Div 3.5 V 3.36 V to 3.64 V
200 mV/Div 1.4 V 1.344 V to 1.456 V
100 mV/Div 700 mV 672 mV to 728 mV
50 mV/Div 350 mV 336 mV to 364 mV
20 mV/Div 140 mV 134.4 mV to 145.6 mV
10 mV/Div 70 mV 67.2 mV to 72.8 mV
5mV/Div
1
35 mV 33.2 mV to 36.8 mV
2mV/Div
1, 2
14 mV 12.56 mV to 15.44 mV
1mV/Div
1, 2
7 mV 5.56 mV to 8.44 mV
1
A blocking capacitor is required at this range to reduce noise. See “Use a Blocking Capacitor to
Reduce Noise" on page 39.
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range. Full scale on all other
ranges is defined as 8 divisions times the V/div setting.