Testing Performance 2
2000/3000 X-Series Oscilloscopes Service Guide 37
c Press the [Acquire] key.
d Then press the Acq Mode softkey and select Averaging.
e Then press the #Avgs softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4 Press the [Cursors] key, set the Mode softkey to Normal, then press the XY softkey
and select Y. Press the Y1 softkey, then use the Entry knob (labeled on the
front panel) to set the Y1 cursor on the baseline of the signal.
5 Use the BNC tee and cables to connect the oscilloscope calibrator /power
supply to both the oscilloscope and the multimeter (see Figure 5).
Table 12 Settings Used to Verify Dual Cursor Accuracy, 3000 X-Series Models
Volts/Div Setting Oscilloscope Calibrator
Setting
Test Limits
5 V/Div 35 V 34.0 V to 36.0 V
2 V/Div 14 V 13.6 V to 14.4 V
1V/Div 7V 6.8V to 7.2V
500 mV/Div 3.5 V 3.4 V to 3.6 V
200 mV/Div 1.4 V 1.36 V to 1.44 V
100 mV/Div 700 mV 680 mV to 720 mV
50 mV/Div 350 mV 340 mV to 360 mV
20 mV/Div 140 mV 136 mV to 144 mV
10 mV/Div 70 mV 68 mV to 72 mV
5mV/Div
1
35 mV 34 mV to 36 mV
2mV/Div
1, 2
14 mV 13.2 mV to 14.8 mV
1mV/Div
1, 2
7mV 6.2mV to 7.8mV
1
A blocking capacitor is required at this range to reduce noise. See “Use a Blocking Capacitor to
Reduce Noise" on page 39.
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range. Full scale on all other
ranges is defined as 8 divisions times the V/div setting.