Testing Performance 2
2000/3000 X-Series Oscilloscopes Service Guide 53
Keysight 2000 X-Series Oscilloscopes Performance Test Record
Serial No. ______________________________________ Test by _____________________________
Test Interval ____________________________________ Work Order No. ______________________
Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0
Accuracy Test 5 V - 250 mV 4.750 V ________
(100 mV + 3% of
threshold setting)
5 V + 250 mV 5.250 V ________
-5 V - 250 mV -5.250 V ________
-5 V + 250 mV -4.750 V ________
0 V - 100 mV -100 mV ________
0 V + 100 mV 100 mV ________
DC Vertical Gain Accuracy
Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
5 V/Div 35 V 33.8 V to 36.2 V ________ ________ ________ ________
2 V/Div 14 V 13.52 V to 14.48 V ________ ________ ________ ________
1 V/Div 7 V 6.76 V to 7.24 V ________ ________ ________ ________
500 mV/Div 3.5 V 3.38 V to 3.62 V ________ ________ ________ ________
200 mV/Div 1.4 V 1.352 V to 1.448 V ________ ________ ________ ________
100 mV/Div 700 mV 676 mV to 724 mV ________ ________ ________ ________
50 mV/Div 350 mV 338 mV to 362 mV ________ ________ ________ ________
20 mV/Div 140 mV 135.2 mV to 144.8 mV ________ ________ ________ ________
10 mV/Div 70 mV 67.6 mV to 72.4 mV ________ ________ ________ ________
5 mV/Div 35 mV 33.4 mV to 36.6 mV ________ ________ ________ ________
2 mV/Div 14 mV 12.72 mV to 15.28 mV ________ ________ ________ ________
1 mV/Div 7 mV 5.72 mV to 8.28 mV ________ ________ ________ ________
Continued on next page.