CAUTION, VENTILATION REQUIREMENTS: When installing the instrument(s) into a
cabinet consideration shall be given to the convection flow into and out of the cabinet.
Consideration shall also be given to the individual instruments to avoid having the
heated discharge of one instrument, now becoming the cooling intake air for another
instrument.
Another area of concern is verification that the maximum ambient operating
temperature of the instrument(s) is not exceeded by cabinet installation.
Keysight recommends forced air convection whenever an instrument(s) are installed in
a cabinet and further recommends that the maximum operating temperature of the
cabinet be reduced 10°C from the lowest, of the maximum operating temperature of a
single instrument.
If there are any concerns or special requirements an Keysight Field Engineer should be
consulted to assure instrument(s) temperature compliance and performance.