Keysight N774-C & 8162-C Series User’s Guide 65
Definition of Terms 3
Measurement: As for "Drift (dark)" (see, Drift (dark) on page 62).
From the measurement result P
i
and the fitted curve F
i
the noise N is
calculated by
N = 2*StDev
i
{P
i
- F
i
}, where StDev
i
{} denotes the standard deviation over
indices i.
Operating humidity
Humidity range where the instrument is designed to be operated.
The instrument must not be operated outside this range. If previously
stored beyond this range, wait for acclimation before turning on the
instrument.
Operating temperature
Temperature range for which the specifications apply if not differently
stated.
The instrument must not be operated outside this range. If previously
stored beyond this range, wait for acclimation before turning on the
instrument.
Operating temperature and humidity
The ambient temperature range and humidity range of the instrument for
which the specifications apply.
Polarization dependent responsivity (PDR)
Specifies the dependence of the measured power on the input state of
polarization (SOP) as ± half the difference between the highest and the
lowest measured power over all SOPs, expressed in dB.
Powers expressed in dBm.
Conditions: As specified.
NOTE
If the instrument is rack- mounted the temperature and humidity within
the rack apply.

Table of Contents

Other manuals for Keysight N774 C Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N774 C Series and is the answer not in the manual?

Keysight N774 C Series Specifications

General IconGeneral
BrandKeysight
ModelN774 C Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals