Keysight N9010B EXA Signal Analyzer Service Guide 221
RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
Connect A13J7 output to a functioning spectrum analyzer and verify the
322.5 MHz intermediate frequency is measuring 32 ± 4 dB as shown in Figure
5-8.
Figure 5-8 322.5 MHz Intermediate Frequency
If this power level is correct the Front End assembly is operating correctly in
high band. If this power level is incorrect verify the following assemblies using
the 4.8 GHz internal calibrator signal.
1. A11 Low Band Switch
2. A10 Input Attenuator B
3. A9 Input Attenuator A
4. A16 Reference Assembly (4.8 GHz calibrator)
5. A12 YTF Preselector
6. A13 Front End Assembly

Table of Contents

Other manuals for Keysight N9010B