Keysight N9010B EXA Signal Analyzer Service Guide 469
Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
9. Refer to Figure 15-25. Connect the two wire harness connectors W30 to
the switches (1). Once connected, position the connections along side the
switches.
Figure 15-25 Wire Harness Routing
Even though the wire colors on W30 harness are different, it does not
matter which one plugs into which switch.