28
Overview
Service Equipment You Will Need
Power Meter Dual Channel
Instrumentation Accuracy: ± 0.5%
Power Reference Accuracy: ± 0.6%
Compatible with 8480 series power sensors
dB relative mode
N1914A E4419A/B
N1912A
A, P
RF Power Sensor
(2 required)
Frequency Range: 100 kHz to 3.6 GHz
Amplitude Range: 30 to +20 dBm
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
VSWR
100 kHz to 1 MHz: 1.20:1
1 MHz to 2 GHz: 1.10:1
2 GHz to 3 GHz: 1.30:1
Cal Factor Uncertainty (std): < 1.6%
Option H84 Cal Factor: Characterized by
standards lab to: ± 0.6%
e
Input Connector: Type-N (m)
8482A
(Option H84)
N8482A
f
N8482A CFT
f
8482A
f
(Non-H84
sensors will
increase
measurement
uncertainty)
A, P
Microwave Power Sensor
(2 required)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: -30 to +20 dB
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
N8485A
f
N8485A CFT
f
8485A
f
A, P
Low Power Microwave
Power Sensor
(2 required)
(for Option P08, P13,
or P26 only)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: 70 to +20 dBm
Zero Set: ± 50 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
8485D
f
A, P
Power Sensor Cable
(2 required)
11730A A, P, T
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Model
a
Alternative
Model
Use
b