EasyManua.ls Logo

Keysight N9020A - Page 29

Keysight N9020A
642 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
Overview
Service Equipment You Will Need
Standards
Frequency Standard Frequency: 10 MHz
Accuracy: < ±1 e10
10
Symmetricom
5071A
Keysight 5061B,
5071A
A, P
50 MHz,
25 dBm Calibrator
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: -45 dBc
Z5602A
Opt H51
A
Attenuators
10 dB
Step Attenuator
Range: 0 to 80 dB
Accuracy:
Characterized by standards lab
g
Calibrated at 50 MHz
VSWR: at 50 MHz: 1.05:1
8496G
(Option 001,
H50)
8496H
(Option 001, H50)
P
1dB
Step Attenuator
Range: 0 to 9 dB
Accuracy:
Characterized by standards lab
g
VSWR: at 50 MHz: 1.05:1
8494G
(Option 001,
H50)
8494H
(Option 001, H50)
P
Attenuator
Interconnect Kit
Type N connector kit to connect 8496G to
8494G
11716A P
Attenuator Driver Compatible with the 8496G and 8494G step
attenuators
11713B 11713A P
3dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 3 dB (nominal)
VSWR: at 50 MHz: 1.2:1
Connector: Type-N (m, f)
8491A
(Option 003)
8491B
(Option 003)
P
6dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: 1.05:1
Connector: Type-N (m, f)
8491A
(Option 006,
H33
h
)
8491B
(Option 006,
H33
h
)
P
6dB
Fixed Attenuator
(2 required)
Frequency: 50 MHz to 3.6 GHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: 1.10:1
Connector: 3.5 mm (m, f)
8493C
(Option 006)
A
10 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 10 dB (nominal)
VSWR: at 50 MHz: 1.05:1
Connector: Type-N (m, f)
8491A
(Option 010,
H33
h
)
8491B
(Option 010,
H33
h
)
P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Model
a
Alternative
Model
Use
b

Table of Contents

Other manuals for Keysight N9020A

Related product manuals