29
Overview
Service Equipment You Will Need
Standards
Frequency Standard Frequency: 10 MHz
Accuracy: < ±1 e10
−10
Symmetricom
5071A
Keysight 5061B,
5071A
A, P
50 MHz,
−25 dBm Calibrator
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: -45 dBc
Z5602A
Opt H51
A
Attenuators
10 dB
Step Attenuator
Range: 0 to 80 dB
Accuracy:
Characterized by standards lab
g
Calibrated at 50 MHz
VSWR: at 50 MHz: ≤ 1.05:1
8496G
(Option 001,
H50)
8496H
(Option 001, H50)
P
1dB
Step Attenuator
Range: 0 to 9 dB
Accuracy:
Characterized by standards lab
g
VSWR: at 50 MHz: ≤ 1.05:1
8494G
(Option 001,
H50)
8494H
(Option 001, H50)
P
Attenuator
Interconnect Kit
Type N connector kit to connect 8496G to
8494G
11716A P
Attenuator Driver Compatible with the 8496G and 8494G step
attenuators
11713B 11713A P
3dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 3 dB (nominal)
VSWR: at 50 MHz: ≤ 1.2:1
Connector: Type-N (m, f)
8491A
(Option 003)
8491B
(Option 003)
P
6dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: ≤ 1.05:1
Connector: Type-N (m, f)
8491A
(Option 006,
H33
h
)
8491B
(Option 006,
H33
h
)
P
6dB
Fixed Attenuator
(2 required)
Frequency: 50 MHz to 3.6 GHz
Loss: 6 dB (nominal)
VSWR: at 50 MHz: ≤ 1.10:1
Connector: 3.5 mm (m, f)
8493C
(Option 006)
A
10 dB
Fixed Attenuator
Frequency: 50 MHz
Loss: 10 dB (nominal)
VSWR: at 50 MHz: ≤ 1.05:1
Connector: Type-N (m, f)
8491A
(Option 010,
H33
h
)
8491B
(Option 010,
H33
h
)
P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Model
a
Alternative
Model
Use
b