550
Assembly Replacement Procedures
Front Frame Assembly
Inverter Board
Removal
1. Refer to Figure 14-89. Using a ¼-inch socket remove the two Hex screws
(1) that secure the flex ribbon into the connector.
NOTE The flex ribbon cannot have tight radial bends as this will damage the cable.
2. Remove the eight screws (2) that secure the inverter top shield to the
bottom shield. Carefully remove the top shield with the ribbon cable
attached
Figure 14-89 Inverter Top Shield Removal