Command Reference
SENSe Subsystem
134
[:SENSe]:TOI Subsection
Set the Average
[:SENSe]:TOI:AVERage:COUNt <integer>
[:SENSe]:TOI:AVERage:COUNt?
[:SENSe]:TOI:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOI:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep. Setting Avg
Number to OFF disables measurement averaging.
*RST: 10/Off
Range: 1 to 1000
Key access:
> Occupied BW > Avg Number On Off
Set the Average Mode
[:SENSe]:TOI:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:TOI:AVERage:TCONrol?
Press Avg Mode to select the type of termination control used for
the averaging function to either Exp (Exponential Averaging mode)
or Repeat. This determines the averaging action after the specified
number of measurements (average count) is reached.
EXP — When you set Avg Mode to Exp, each successive data
acquisition after the average count is reached is exponentially
weighted and combined with the existing average. Exponential
averaging weights new data more than old data, which
facilitates tracking of slow changed signals. The average is
displayed at the end of each sweep.
Repeat — When you set Avg Mode to Repeat, after reaching the
average count, all previous result data is cleared and the
average count is set back to 1.
*RST: EXPonential
Example: TOI:AVG:TCON EXP
Key access:
> More > TOI > Avg Mode
Meas