Command Reference
SENSe Subsystem
[:SENSe]:DETector Subsection
Automatic Detection Type Selected
[:SENSe]:DETector:AUTO OFF|ON|0|1
[:SENSe]:DETector:AUTO?
Switches automatically to the optimum detection type for typical
measurements using the current instrument settings.
The detector type is average if any of these are on:
Trace averaging when the Average Type is Power (RMS).
The detector type is sample if any of the following conditions are
true:
Trace averaging is on with average type of video
Both max and min hold trace modes are on
Resolution bandwidth is less than 1 kHz, and trace averaging is
on
The detector type is negative peak if any trace is in min hold and no
traces are in max hold.
The detector type is peak if the above conditions are off.
Manually changing the detector function turns Auto off.
Refer to Figure 20 on User’s Guide, which shows a decision tree of
how detection type is determined.
*RST: On
Key access: >
Detector
Type of Detection
[:SENSe]:DETector[:FUNCtion]NEGative|POSitive|SAM-
Ple|AVERage|RMS|NORMAL
[:SENSe]:DETector[:FUNCtion]?
Specifies the detection mode. For each trace interval (bucket),
average detection displays the average of all the samples within
the interval. The averaging can be done using two methods:
the power method (RMS)
the video method (Y Axis Units)
Det/
Display